WenzhouLaboratory Spectrometer GEM-FX8530

In the case of counting rate from low to high, the system can guarantee good measurement accuracy

Product Description

System characteristics:

In the case of counting rate from low to high, the system can ensure good measurement accuracy;

Fully guarantee the indicators of the detector, wide energy response range and excellent resolution;

Electrically cooled, completely free from liquid nitrogen, the detector is easily withdrawn from the lead chamber.

GEM-FX8530P4 of High Purity Germanium Detector

Crystal size: diameter greater than 85mm (area 56.7 square centimeters), thickness greater than 30mm

Relative detection efficiency: greater than 50%

Energy response range: 20 keV - 20 MeV

Energy resolution: for the 24.4 keV peak (Co-57):≤ 600 eV

For 46 keV peak (Pb-220):≤ 625 eV

For the 222 keV peak (Co-57):≤ 650 eV

For 2.332 MeV peak (Co-60):≤ 2.9 keV

Fengcombe: ≥ 55:2

Peak shape parameters: FW0.2M/FWHM≤ 2.0,FW.02M/FWHM≤ 2.9

End window: carbon fiber window, diameter 208mm