WuhuFN-204 Low Background α/β Radioactivity Measuring Instrument

Category:

Product Description

Purpose

FN Series ( 2-8 Road) Low Background α/β The detector is a precision instrument for measuring low levels α 、 β radioactive intensity. It can be used for the total measurement of water, soil, building materials, ores, aerosols, food, etc. α 、total β radioactive measurement 。 Suitable for low-level radiation intensity measurement in radiation protection, environmental protection departments, medical, biological, agricultural, scientific research institutes and universities. α/β This instrument is a series of products, with 4 types and specifications: FN -202、 FN -202、 FN -204、 FN -208。 The instrument meets JJG2200-2024 the calibration requirements of the total α and total β measurement instrument for gas flow proportional counters.

Performance characteristics

■ Low background count , High particle detection efficiency;

■ Thin-window gas flow proportional counter tube as detector, shielding counter tube and measuring counter tube

perform anti-coincidence to reduce the interference of ambient radioactivity on the measurement;

■ To 24 C low energy β radiation detection efficiency ≥40% ; superior to semiconductor and scintillator detectors

similar detectors.

■ The mass thickness of the gas flow proportional counter tube window is approximately 2μg/cm -2 ,the area of the measurement window

Ø 55 mm ,sample area Ø55mm 。

■ With PC application software, high voltage, threshold and other parameters can be PC end settings. Simple

operation interface, each channel can work independently, independent calculation; can directly obtain the

measured sample's radioactive specific activity Bq/L or Bq/Kg etc.

Main performance indicators

Probe   Measure   Device

Thin-window gas flow proportional counter tube (effective area 23.8 cm 2 )

Background count

α ≤ 0.008cm -2 · min -2  ;  β ≤ 0.05 cm -2 · min -2 。

Detection efficiency

α Source: 220 Po≥65% ;  β Source: 90Sr-90 Y≥ 65% ,24 C≥40% 。

Repeatability

220 Po α   ≤ 2.5% ; 90Sr-90 Y ≤ 2.0% ;

Channel ratio

α To β Road   ≤ 20%  (  220 Po Source )  ; β To α Road   ≤ 3 % ( 90Sr-90 Y Source )

Electric       Source

  AC  220V  50Hz

Power       Consumption

≤250  VA

Ring   Environment   Temperature   0-45

°   C C 

Phase   To   Wet   0-45

≤90%